![]() |
|||||||||||
|
|||||||||||
|
About
Absorbance Spectroscopy |
|||||||||||
|
|
|||||||||||
About UV-Vis Molecular Absorbance Spectroscopy In molecular absorbance
spectroscopy a beam of ultraviolet or visible light is directed through a
sample. Some of the light may be transmitted through the sample. Light that
was not transmitted through the sample was absorbed. A molecule can absorb some of the light only if it can accomodate that additional energy by promoting electrons to higher energy levels. The energy of the light being absorbed must match the energy required to promote the electron. Therefore, not all wavelengths of light are absorbed equally by a sample. An absorbance spectrum depicts what wavelengths of light are absorbed by a sample. The UV-Vis absorbance spectrum below was obtained by passing different wavelengths of light through a solution of fabric dye, and measuring the intensity of light (P) passing through the solution. One can readliy see what wavelengths of light are absorbed (peaks), and what wavelenghts of light are transmitted (troughs).
One very important relationship in absorbance spectroscopy is Beer's Law: A=abc where: A is absorbance This relationship is the basis of all quantitative work in absorbance spectroscopy. It allows one to determine the concentration of an absorbing species simply by measuring its absorbance. |
|||||||||||
![]() |
|||||||||||
![]() |
|||||||||||
|
The Cary 50 design is simple, yet elegant. Its fast-scanning monochromator collects spectra in seconds, and the Xe flash lamp is pulsed to avoid interference from room light. The Cary 50 is controlled by intuitive software which allows students who are unfamiliar with spectrophotometers to begin collecting data without extensive software training. We have purchased several accessories for use with this instrumen t that include a peltier thermostatted cell holder to rapidly change sample temperature between 0 to 100 oC, sample temperature probes to accurately determine the temperature of the sample inside the cuvette, and a stopped flow (rapid kinetics) accessory to measure kinetics of fast reactions (less than a second). |
|||||||||||
|
|
|||||||||||
www.thermo.com
|
Genesys 10 UV Spectrophotometers We have a fleet of 8 Genesys 10 UV spectrophotometers that are used in a variety of laboratory courses throughout the chemistry curriculum. These UV-Vis spectrophotometers are rugged, user-friendly, have a small footprint, and are great for quantitative work and kinetics at a single wavelength. |
||||||||||
|
|
|||||||||||
Hewlett Packard 8451A Diode Array Diode array spectrometers contain thousands of individual photodetectors enabling them to measure the intensity at many wavelengths simultaneously. Absorbance spectra can be obtained in a fraction of a second, and several wavelengths of light can be monitored simultaneously during kinetic experiments. This particular diode array spectrometer has an extremely rugged optical bench (only one moving part), and has been in operation at KU since the mid 1980's. |
![]() |
||||||||||
|
|
|||||||||||
Hitachi U-3110 Double Beam Spectrophotometer Double beam spectrophotometers split the light beam into two paths. One path passes through the sample, and the other path passes through a blank. A spectrophotometric blank is basically everything that is in the sample except the component you are trying to measure. With the double beam approach P and Po are measured at the same time (or nearly so), which corrects for a variation in lamp intensity or drift in instrument electronics.This instrument was obtained from a generous donation from Air Products and Chemicals, Inc. |
![]() |
||||||||||
|
|
|||||||||||
Ocean Optics Miniature Spectrophotometers Ocean Optics has taken the lead in miniaturizing diode array spectrometers. By using the diode array approach, there are no moving parts in the spectrometer. Fiber optic cables are used to take light from the source to the sample, and from the sampe to the detector. These are versatile instruments that can be configured to apply to a variety of sampling situations. |
|||||||||||
|
|
|||||||||||
|
Spec 20's have been around since the 1960s, and we have a few of those. These spectrometers are rugged, rudamentary, reasonably priced, visible spectrometers. It is because of these qualities that I am not surprised to see Spec 20s still in use in many educational and industrial labs. The 8 Spec 20s we have are all operational. One of them has undergone a solid-state upgrade, and is fitted with a digital display. We have a box of spare vacuum tubes, light bulbs, and detectors to keep them going well into the next decade. |
|||||||||||
![]() |
![]() |
||||||||||
|
|
|||||||||||
Applications of UV-Vis Molecular Absorbance Spectroscopy Quantitative analysis
of all molecules that absorb light |
|||||||||||
|
|
|||||||||||
Links to Molecular UV-Vis Sites Several Vendors of UV-Vis Spectrophotometers Varian,
Inc. Agilent Ocean
Optics Hitachi
High-Technologies UV-Vis Resources Basic
UV-Vis Theory, Concepts and Applications Bunsen
Learner (University of Edinburgh) |
|||||||||||